Maintain the instrument
■
Backup or restore the instrument ....................................... 70
■
Decontaminate the sample block ....................................... 71
■
Replace the instrument fuses ........................................... 73
■
Prepare the instrument to store, move, or ship ........................... 74
IMPORTANT! This chapter contains user maintenance procedures for the instrument.
Procedures other than those described in this document must be performed by a
qualied Thermo Fisher Scientic service representative.
Backup or restore the instrument
In the home screen, touch Seings4Maintenance and Service4Backup/Restore.
To Action
Backup
1. Touch Backup Instrument.
2. Select a location to store the backup settings for your
instrument.
3. If backing up to a USB, insert a USB into the front-panel USB
port.
4. Enter a backup file name, then touch Done.
5. Select the elements to backup, or leave them all selected.
6. Touch Backup.
Restore
(Administrator only)
1. Touch Restore a Backup.
2. Select the storage location of the backup settings.
3. If restoring from a USB, insert a USB into the front-panel
USB port.
4. Select the backup file, then touch Restore.
6
70
QuantStudio
™
1 Real-Time PCR System Installation, Use, and Maintenance Guide