EasyManua.ls Logo

Thermo Scientific QuantStudio 1 - Page 69

Thermo Scientific QuantStudio 1
108 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8.
Review the plot. Passing calibration results show uniform signals with peaks that
are aligned with the dye wavelength.
Figure 3 Example dye calibration plot
Note: The peaks for your dye may align with a different filter set.
9.
Select an action depending on whether the custom dye calibration passed or
failed.
Calibration status Action
Passed
Touch Accept Results or Reject Results.
Note: Accepting the results saves the calibration data to
the instrument and overwrites existing data.
(
Optional
) Touch Transfer EDS to transfer the calibration
data to a USB.
Failed
Perform the calibration again using a new custom dye
plate. Create the plate using the next dye concentration
greater than the optimal dye concentration determined in
“Use a dilution series to determine an optimal custom dye
concentration“ on page 62
See “Troubleshoot calibration failure“ on page 55.
10.
Unload the plate from the instrument.
Note:
·
You must also add the custom dye to the QuantStudio
Design and Analysis
Desktop Software dye library before creating, running or analyzing experiments
that use custom dyes.
·
To perform a custom melt experiment, you can either create a Standard Curve or a
Custom experiment with melt, then specify the data points per degree in the
method.
Chapter 5 Calibrate and verify instrument performance
Calibrate for a custom melt curve experiment
5
QuantStudio
1 Real-Time PCR System Installation, Use, and Maintenance Guide
69

Table of Contents

Related product manuals