Calibrate custom dyes
The instrument can run assays designed with custom dyes. Custom dyes include:
Dyes that are not manufactured by Thermo Fisher Scientic.
Dyes or formulations of dyes that are not system dyes for the instrument.
For custom dye requirements, see “Custom dyes“ on page 10.
Workflow: Custom dye calibration
For each custom dye, determine the optimal dye concentration. Use this concentration
to prepare all subsequent dye calibration plates.
Use a dilution series to determine an optimal custom dye concentration
“Prepare a custom dye dilution plate“ on page 63
“Run the dilution plate as an experiment“ on page 64
“Determine the optimal dye concentration“ on page 64
Calibrate the custom dye using the optimal concentration
Create a custom dye calibration plate
Add a custom dye to the instrument
“Perform a custom dye calibration“ on page 66
Custom dye dilution guidelines
Prepare a dilution series for each custom dye.
Target several dye concentrations within a range of 100–2,000 nM.
Choose a 2- or 3-fold dierence in dilution points.
Dispense 10–20 µL per well in a 96-well plate.
Dilute the dye in buer compatible with your master mix.
(Intercalating dyes only) Add the appropriate amount of amplied PCR product to
generate uorescence.