The Ethernet port of the instrument supports:
• Static IP network service with subnet mask, primary and secondary data network
service (DNS), and default gateway seings, or dynamic host conguration
protocol (DHCP) network service.
• mDNS/DNS for local domains.
Note: Because mDNS is limited to direct network connections, an instrument
congured for mDNS may not be visible to other nodes that are separated by a
router, hub, or another network device.
• IPv4 linknlocal (IPV4LL) in the RFC (also known as Automatic Private IP
Addressing [APIPA] or Internet Protocol Automatic Conguration [IPAC]).
Note: When an instrument is set for DHCP, APIPA is automatically enabled, and
the instrument provides an IP address when no address is supplied by the DHCP
server.
Ports
Condition
80/443
Standard ports for
instrument-to-Cloud and computer-to-Cloud connections
mDNS, 7000 Instrument-to-computer connection
mDNS, 5353 Instrument discovery
• Consult a network administrator before connecting the instrument to a network.
• To enable the full functionality of the software, the computer requires a network
connection.
• Open the rewall port for the instruments to be discovered. See “Firewall ports
that must be open“ on page 83.
• Observe the restrictions to mDNS and Autodiscovery.
The instrument supports mDNS but only when the instrument and computer
share a direct network connection and are within the same subnet. Network
computers that are separated from the instrument by a router, hub, or another
network device may not be able to access the instrument by its host name.
• Conrm the uniqueness of the instrument name.
– The instrument name must be unique within the subnet. The desktop
software can automatically discover instruments on the link-local network.
– The instrument does not test the uniqueness of the instrument name within
the subnet when it is set.
Ethernet port
overview
Firewall ports that
must be open
Networking
guidelines and
best practices
Appendix A Install and connect the instrument to a network
Networking
A
QuantStudio
™
1 Real-Time PCR System Installation, Use, and Maintenance Guide
83