Calibration description and purpose
Pass Criteria
Dye
The software extracts a spectral profile for each dye standard, then
produces a set of spectral profiles plotted as fluorescence vs filter.
The software uses calibration data to characterize and distinguish the
individual contribution of each dye in the total fluorescence signals
collected by the instrument.
Dye spectra peak within the same
filter as their group.
View the calibration status and set reminders
1.
In the home screen, touch Seings4Maintenance and
Service4Calibrations4History and Reminders.
2.
In the Calibration Reminders screen, view the status of each calibration type.
3.
(Optional) Touch a calibration row to view the history of that specic calibration
type, then touch Done.
4.
Touch Edit to set the calibration reminder seings. For each calibration type:
a.
Slide the control On to enable the calibration reminder.
b.
Edit the Exp interval and Remind me elds.
c.
Touch Save.
5.
(Optional) To transfer the calibration report, touch Export then follow the
directions on the screen.
6.
Touch Done.
Note: The calibration reminders feature requires a connection between the
instrument and a computer network.
1.
In the Thermo Fisher Cloud, click
for the Instrument Connect page.
2.
Select any of your registered instruments.
3.
In the Summary tab:
Click Calibrations to view the status of each calibration type.
(Optional) Click Calibration Reminders to set the calibration reminder
time table and enter the notication email address(es).
Note: More than one email address can receive the calibration reminders.
Note: The seings are automatically saved.
(Optional) In the Downloads section, click Maintenance Summary.pdf to
download the calibration status report.
View calibration
status and set
reminders in the
instrument
View calibration
status and set
reminders in the
Cloud
Chapter 5 Calibrate and verify instrument performance
View the calibration status and set reminders
5
QuantStudio
1 Real-Time PCR System Installation, Use, and Maintenance Guide
51

Table of Contents