Template and
experiment
components
Instrument Desktop Cloud
Plate Sample names
You cannot edit targets/SNP assays
or tasks on the instrument.
Define and assign samples, targets or SNP assays, and
tasks in the Quick Setup and Advanced Setup panes of the
Plate tab.
Run Start and monitor a run in progress
View: Time remaining and
Temperature, Method, Plots
Pause, Resume, Stop a run
Start and monitor a run in
progress
View: Time remaining and
Temperature, Method, Plots
Monitor a run in progress
(link to Thermo Fisher Cloud
Instrument Details)
View: Time remaining and
Temperature, Method, Plots
Results Not applicable Review plots
Export Not applicable Export results
Before installing third-party software on the computer running QuantStudio
™
Design
and Analysis Desktop Software, conrm that the third-party software will not do the
following:
• Restrict Ethernet communication.
• Interfere with instrument or computer operation.
Supported options for instrument and computer connections
We support the following direct, networked (LAN– local area network), or Cloud
congurations. Congurations other than those listed are not recommended. Select a
conguration that meets the needs of your laboratory's instrument, software, and
workow requirements.
IMPORTANT! Do not connect both an Ethernet cable and the Wi-Fi module
(Cat. No. A26774) to the instrument. Conguring the instrument for both wired and
wireless connection can interfere with instrument operation.
Third-party
software
Chapter 1 Product information
Supported options for instrument and computer connections
1
16
QuantStudio
™