EasyManua.ls Logo

Thermo Scientific QuantStudio 1 - Perform Rnase P Verification

Thermo Scientific QuantStudio 1
108 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Thaw, vortex, and centrifuge an RNase P plate
IMPORTANT! Expose the RNase P plate to room temperature for no more than
45 minutes, inclusive of thawing and preparation time.
After thawing, the RNase P plate cannot be refrozen.
1.
Remove the RNase P plate from the freezer, then thaw the plate in its packaging.
Thaw the plate for approximately 5 minutes.
Use the plate within 30 minutes of thawing.
2.
Conrm that the bench, vortex, and centrifuge are clean. Before use, wipe the
vortex and centrifuge using a lint-free tissue.
3.
While wearing powder-free gloves, remove the plate from its packaging.
4.
Vortex the plate for 5 seconds, then centrifuge at 750–1,000 × g for 2 minutes.
5.
Conrm that the liquid in each well is at the boom of the well and free of
bubbles. If it is not, centrifuge the plate again.
IMPORTANT! Keep the boom of the plate clean. Fluids and other contaminants
on the boom of the plate can contaminate the sample block and cause an
abnormally high background signal.
1.
In the home screen, touch Seings4Maintenance and Service4RNase P
Verication.
2.
Load the plate into the instrument.
3.
Touch Start.
4.
When the run is complete and the screen displays Verication Complete, touch
View Results to conrm the status of the run.
Calibration status Action
Passed Instrument is ready for use.
Failed See “Troubleshoot verification failure“ on page 61.
Perform RNase P
verification
Chapter 5 Calibrate and verify instrument performance
Perform instrument verification using RNase P plates
5
QuantStudio
1 Real-Time PCR System Installation, Use, and Maintenance Guide
59

Table of Contents

Related product manuals