5.
In the RNase P Verication Status screen, touch one of the following options:
• Accept Results — Save the results to the instrument.
• Reject Results — Delete the RNase P verication results.
• Export Results — Export the calibration results to a USB.
6.
Unload the plate from the instrument.
CAUTION! PHYSICAL INJURY HAZARD. During instrument operation,
the plate temperature can reach 100°C. Allow it to cool to room
temperature before handling.
Chapter 5 Calibrate and verify instrument performance
Perform instrument verification using RNase P plates
5
60
QuantStudio
™
1 Real-Time PCR System Installation, Use, and Maintenance Guide