EasyManua.ls Logo

TRUE SOLTEC CFM-BL10Pro - CFM Cannot Detect Subtle Defective Samples

TRUE SOLTEC CFM-BL10Pro
87 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CFM-BL10Pro General instruction manual
80 / 87
Cause 3 Area divisions of the T1/T2/T3 are not correct and CFM cannot recognize
differences between good and bad samples.
As you can see the figure below, if the judgement area division is not correct, CFM
cannot detect defective samples.
The initial division lines are T1 Start5%, T2 Start30%, T3 Start70% and T3 End
90%. Please refer to 5.2.1 Program edit screen for the details of the Area.
If you cannot improve the accuracy of CFM judgement, revise the tolerance setting
as well.
8.2. CFM cannot detect subtle defective samples.
Detection of one strand cut and very subtle insulation crimp is affected by
combinations of terminals and wires, press machine or applicator conditions. With a
severe tolerance, CFM detects many kinds of defective samples, but many false
judgements occur. Crimp the defective samples which you want to detect and find the
tolerance that CFM can detect the defective samples with. Then check if the false
judgements occurred with the tolerance are acceptable.
T1 Start
T2 Start
T3 Start
T3 Start
T1
T2
T3
Good area division
T1
T2
T3
Wrong area division

Table of Contents

Related product manuals