CFM-BL10Pro General instruction manual
80 / 87
Cause 3 : Area divisions of the T1/T2/T3 are not correct and CFM cannot recognize
differences between good and bad samples.
As you can see the figure below, if the judgement area division is not correct, CFM
cannot detect defective samples.
The initial division lines are T1 Start:5%, T2 Start:30%, T3 Start:70% and T3 End:
90%. Please refer to 5.2.1 Program edit screen for the details of the Area.
If you cannot improve the accuracy of CFM judgement, revise the tolerance setting
as well.
8.2. CFM cannot detect subtle defective samples.
Detection of one strand cut and very subtle insulation crimp is affected by
combinations of terminals and wires, press machine or applicator conditions. With a
severe tolerance, CFM detects many kinds of defective samples, but many false
judgements occur. Crimp the defective samples which you want to detect and find the
tolerance that CFM can detect the defective samples with. Then check if the false
judgements occurred with the tolerance are acceptable.
T2 Start
T3 Start
T2
T2