104 ACF5000 FTIR ANALYZER SYSTEM | OI/ACF5000-EN REV. A
Change number of decimal places
Menu path
MENU → Configure → Component-specific → Measuring Range
(→ Select component) → ...
Selection
Displayed are all measuring ranges configured (at the plant) for a measured
component.
Procedure
Select measuring range with the arrow keys, press DECIMAL PLACES, set
the number of decimal places with the arrow keys and confirm with ENTER.
NOTICE
The setting only affects the display of the measured values in the display.
The number of decimal places cannot be increased.
The changed number of decimal places is displayed after switching in meas-
uring mode in the display.
Number of decimal places
For the display of the measured value in physical units (e.g. ppm), the num-
ber of decimal places depends on how large the span of the measuring
range is:
Measuring span Decimal places
≤ 0.05 5
≤ 0.5 4
≤ 5 3
≤ 50 2
≤ 500 1
> 500 0
For the display of the measured value in % of the measuring range scope
(%Span), two decimal places are always displayed.
The number of decimal places when setting the parameters is the same as in
the display in measuring mode.