R3132 Series Spectrum Analyzer Operation Manual
3.3.18 POWER MEASURE Key (Power Measurement)
3-60
SEPA: The screen mode is changed to SEPA when a sweep
takes place.
CARRIER:
The entire screen is used.
Calculates the ACP based on the power along the band
specified by Carrier Band Width.
Parameter Setup Displays the Prmtr Setup menu.
Default Sets the following settings to default values: the frequency span,
resolution bandwidth, video bandwidth, sweep time, detector,
trace mode, CS/BS Setup and Nyquist Filter Setup. When De-
fault is enabled, the defaults are automatically set when the ACP
is measured.
Manual Disables the Default function. As a result, the above settings must
be set manually. The currently set values are used when the ACP
is measured.
Define Default The currently set values become the new defaults.
Graph Displays the Multi MKR menu (used for ACP Graph).
Turns the graphic display function on.
Displays the leakage power graph as Trace B and puts Delta
marker in the center of the screen.
The B memory is used to display the ACP graph.
Marker No. Enter the multi marker number here.
Marker ON Displays the multi-marker specified at Marker No. in the center of
the trace and the frequency and level of the marker in the marker
area.
Marker OFF Removes the multi-marker specified by the number.
Active Marker Turns on or off the multi-marker displayed on the screen.
Reset Marker Removes all multi-markers except for multi-marker 1.
Graph OFF Turns the graphic display function off.
ACP OFF Turns off the ACP measurement function, and returns to the pow-
er menu.
1/2 more Displays the Power menu (2/2).
Spectrum Mask Displays the Spect Mask menu.
Displays the measurement and judgment results in combination
with the pass/fail judgment result using the limit lines for the ACP
measurement.