R3132 Series Spectrum Analyzer Operation Manual
4.2.9 GPIB Command Codes
4-36
Table 4-11 Meas (2 of 3)
Function Listener Code
Talker Request
Code Output Format
Phase noise
measurement
C/N measurement mode
ON CNIS ON CNISON? 0: OFF
OFF CNIS OFF 1: ON
Offset frequency data read-
out
––– CNIS? n<DLM> f1, l1 <DLM>...(*1)
Table input CNOFSIN * ––– Offset frequency
Deleting the table CNOFSDEL ––– ––
Signal track ON CNSIG ON CNSIG? 0: OFF
OFF CNSIG OFF 1: ON
Average number CNAVG [ON,] *
CNAVG OFF
CNAVG?
CNAVGON?
Integer (2 to 999)
0: OFF
1: ON
Phase jitter
measurement
Phase jitter measurement
mode
ON PJIT ON PJITON? 0: OFF
OFF PJIT OFF 1: ON
Result value readout –– PJIT? Carrier level, total SSB noise and phase jitter
Start offset frequency PJSRTO * PJSRTO? Offset frequency
Stop offset frequency PJSTPO * PJSTPO? Offset frequency
Signal track ON PJSIG ON PJSIG? 0: OFF
OFF PJSIG OFF 1: ON
Average number PJAVG [ON,] *
PJAVG OFF
PJAVG?
PJAVGON?
Integer (2 to 999)
0: OFF
1: ON
(*1) n=Set number
fn=Offset frequency
ln=Level
<DLM>=Delimiter

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ADVANTEST R3162 and is the answer not in the manual?

ADVANTEST R3162 Specifications

General IconGeneral
BrandADVANTEST
ModelR3162
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals