R3132 Series Spectrum Analyzer Operation Manual
5.4.1 Offset Error (Internal Mixer Mode)
5-110
2. Set the signal generator as follows:
Frequency: 30 MHz
Output level: 0 dBm
3. Set the R3132 series as shown below after the preset:
Center frequency: 30 MHz
Frequency offset: -30 MHz
4. Perform an FM Demod calibration for the R3132 series as shown below:
UTILITY, FM Demod, Demod Cal and All.
5. Press the RETURN key after the calibration finishes.
6. Set VBW used for the R3132 series to 10 Hz.
7. Press the MKR key to display the marker.
8. Set the frequency range as shown below.
UTILITY and Range.
9. Sets the frequency range listed in the performance verification sheet.
10. Press the SINGLE key to read the marker frequency in absolute value, and
record it on the verification sheet.
11. Repeat steps 9 through 10 for each frequency range.