R3132 Series Spectrum Analyzer Operation Manual
6.3.1 External Mixer Performance Verification Test Procedures
6-33
Part 2 Frequency Range 70 GHz to 80 GHz
17. On the R3172/R3182, press keys as follows to preset.
SHIFT, CONFIG (PRESET)
18. Insert the correction data floppy disk.
19. On the R3172/R3182, press keys as follows to set FD reading.
RECALL, Device, RAM/FD (FD)
20. Select the WHM15-2 file by rotating the data knob.
21. On the R3172/R3182, press RECALL key to load the correction data.
22. On the R3172/R3182, set controls as follows;
Frequency Span: Zero
Reference Level: -40 dBm
dB/div: 10 dB/div
RBW: 1 kHz
VBW: 10 Hz
Detector: Sample
23. On the R3172/R3182, set control as follow;
Center Frequency: 70 GHz
24. On the R3172/R3182, press keys as follows to set number of average to 3 times.
TRACE, 1/2_more, AVG A, 3, Hz (ENTER)
Wait until the average process has completed
25. On the R3172/R3182, press PKSRCH to capture the highest noise signal.
26. Record the level of marker on the performance verification test record sheets.
27. On the R3172/R3182, set control as follow;
Center Frequency: 70.5 GHz
Wait until the average process has completed
28. On the R3172/R3182, press PKSRCH key to s capture the highest noise signal.
29. Record the level of marker on the performance verification test record sheets.
30. Repeat steps 23 through 29 for incrementing the center frequency of 500 MHz
step up to 80 GHz.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ADVANTEST R3162 and is the answer not in the manual?

ADVANTEST R3162 Specifications

General IconGeneral
BrandADVANTEST
ModelR3162
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals