Changing Replenishment Rates
Follow this procedure to change the replenishment rate.
Step 1: Press DEV and REPL.
The current developer replenishment rate appears. For
example:
DEV REPL 200 CC
Step 2: Press PRESET.
The preset rate appears. For example:
DEV SET 200 CC
Step 3: Press plus (+) or minus (-) to change the setting.
Step 4: Press ENTER to store the setting that appears on the display.
To adjust fixer replenishment, press FIX instead of DEV in Step1.
Range for Developer and Fixer Replenishment: 10–999 ml.
Avantra 20/25 OLP: Processor Operation
9–15