Vials and Bottles 6
Installation, Operation, and Maintenance 115
Part 2, Operation
10° and crimp it again. Adjust the crimper for a looser crimp by turning the
adjusting screw clockwise.
2 Check that the cap cannot be turned by hand. If the cap is loose, adjust the
crimper for a tighter crimp by turning the adjusting screw
counterclockwise. Crimp the cap again.
3 Be sure that each cap has a flat septum centered over the top of the vial.
If the septum is not flat, remove the cap, turn the crimper adjusting
screw clockwise, and try again.
If the cap is not centered, remove the cap and make sure the new cap is
flat on the top of the vial before you squeeze the crimper.