Maintenance 8
Installation, Operation, and Maintenance 155
Part 3, Maintenance and Troubleshooting
7 Turn the needle support assembly so that the flat surface of the slide glides
up and down the syringe carriage tracks (Figure 66).
8 Align the bearing on the needle support assembly with the plastic bearing
clip to the right of the syringe latch and push gently on the bearing until the
assembly snaps into place.
9 Place the injector back on the GC.
Figure 66 Installing the needle support assembly
Syringe latch
Needle
support
tracks
Cross-section
of slide alignment
Slide
CAUTION
Be careful not to bend the needle during installation.
CAUTION
Do not operate the injector without a syringe or align probe in place because the free
swinging syringe latch may interfere with the motor and jam the syringe carriage.