58 Installation, Operation, and Maintenance
Part 1, Installation
2 Installation
The three-conductor power line cord, when plugged into a properly grounded
receptacle, grounds the instrument and minimizes shock hazard. A properly
grounded receptacle is one that is connected to a suitable earth ground.
Proper receptacle grounding should be verified.
Line voltage
The G2912A ALS controller operates from one of the AC voltage supplies
listed in Table 9, depending on the standard voltage of the country from which
it was ordered. It is designed to work at a specific voltage; make sure your
instruments voltage option is appropriated for your lab. The voltage
requirement is printed near the power cord attachment. See “Verifying
G2912A ALS Controller power configuration” on page 59 for setting the proper
voltage for this controller.
WARNING
Any interruption of the grounding conductor or disconnection of the power cord
could cause a shock that could result in personal injury.
Table 9 Voltage requirement by country
Country Voltage Part number
Americas, 10 amp 120 V 8120-1378
Argentina 220 V 8120-6869
Australia, 10 amp 240 V 8120-1369
Chile, 10 amp 220 V 8120-6978
China, 10 amp 220 V 8121-0723
DK/Greenland, 10 amp 220 V 8120-3997
Europe, 10 amp 230 V 8120-1689
GB/HK/SG/MY, 10 amp 240 V 8120-8705
India/S Africa, 10 amp 240 V 8120-4211
Israel, 10 amp 220 V 8120-5182
Japan, 10 amp 200 V 8120-4753
Korea, 10 amp 220 V 8121-1226
Switzerland, 10 amp 230 V 8120-2104