5. Monitoring Output
119
ML-5120A
Chapter 2 Various Settings
Introduction Part
Installation and
Preparation Part
Operating Part Maintenance Part Appendixes
When the total laser output time is higher than 0.5 s in the CW mode, set “HIGH”
and “LOW” at the ratio of output (PEAK POWER × POWER [%]) set on the
SCHEDULE screen. The following formula is given to convert the set value
into Watts. POWER [%] is also increased or decreased according to the output
increase/decrease between each POINTs. When the measured value of the
monitored laser light is out of the range shown by red lines below, error No.035/
LASER POWER OUT OF RANGE occurs.
Output
Time
Controlled
waveform
PEAK POWER (1) × POWER [%] (2) / 100
× HIGH [%] (3) / 100
PEAK POWER (1) × POWER [%] (2) / 100
× LOW [%] (4) / 100
(1)
(2)
(3)
(4)