User Manual – Rev BE AMETEK Programmable Power
MX Series 290
PAUSE
Sets the time in seconds for which the test will pause between ranges. There is
only one value for the entire test. The PAUSE can be set when the test is not
running. For the Meister curve test, the pause time is normally set to 0 secs. If
a value other than 0 is entered, the interharmonics are set to 0% during the
pause times.
RESONANT
This field is used to report the resonant points, if any, after running the Meister
curve test. No editing is allowed in this field. When the test is completed, a
display of harmonics current versus frequency plot is available. To view the
graph, select the field and press the ENTER key.