User Manual – Rev BE AMETEK Programmable Power
MX Series 291
9.5.4.3 RUNNING THE TEST
The test sequence used is a function of the EUT class. The end user needs to
determine the appropriate class of EUT. The test protocol is document in the IEC
61000-4-13 test standard. For reference, the test flow charts are shown here. It
is assumed that the end user has a copy of the actual test standard, however.
Figure 9-28: IEC 61000-4-13 Test Flowchart Class 1 and 2
Start - Class 1/2
Determine appropriate test
configuration
Perform 8.2.1 Test
"Harmonic combination"
Flatcurve and Overswing
Any functional
anomalies?
No No
Class 2
required?
'Meister
curve' required?
Any functional
anomalies?
Any functional
anomalies?
Any functional
anomalies?
Test Completed
FAIL
Record Results
Test Completed
PASS
Record Results
Yes
Yes
No
Perform 8.2.2 Test
"Frequency Sweep"
Perform 8.2.4 Tes
"Meister Curve"
Perform 8.2.3 Test
"Individual Harmonics/
Interharmonics"
Yes
Yes
No
Yes
Yes
No
No

Table of Contents

Other manuals for Ametek MX Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek MX Series and is the answer not in the manual?

Ametek MX Series Specifications

General IconGeneral
BrandAmetek
ModelMX Series
CategoryPower Supply
LanguageEnglish

Related product manuals