7-24
129390199 Operator Manual Cycle and Control Value Programming
7.5.5 Utility Control with
Integral Electric Steam
Generator
If the sterilizer is set up to be equipped with an integral electric
steam generator, the following screen appears on the display when
Utility Control is accessed. This screen adds the option for setting a
daily generator flush.
• The automated generator flush can only be programmed if the
Steam Generator button on the Configure Machine screen (in
Service Mode) has been set to Yes. Refer to M
AINTENANCE
MANUAL (P764333-449) for Information on this setting. This
setting should only be changed by a trained, authorized service
technician.
• The automated generator flush only occurs following an
automatic utility Shutoff, and once the generator has dropped
below a preset temperature. If the generator does not reach the
preset temperature before automatic utility restart, the flush will
not occur. It is recommended that utility off time be set for at
least six to eight hours, depending on atmospheric conditions.
Ambient temperature conditions may increase or reduce cool-
down time.
• The duration of the generator automated flush is three minutes.
1. If the Generator Flush button displays No, press Generator
Flush button to enable automatic generator flush.
2. If the Generator Flush button displays Yes, press Generator
Flush button again to disable automated generator flush.
3. Press Exit to return to the machine setup screen.