129390199 Operator Manual Sterilizer Operation
Table 6-10. Factory-Set Liquid Cycle and Cycle Values
The AMSCO Evolution Series Prevacuum Sterilizer is equipped with the following factory programmed liquid
sterilization cycle and cycle values.
Cycles Sterilize Temp.
Sterilize
Time
†
Dry
Time Recommended Load
Validation
Standard
LIQUID
*
250°F (121°C) 45 Min. N/A Three 1000 mL flasks.
R
EFER TO Table 6-5 FOR
GUIDELINES.
ST-8
* Liquids sterilized using the Liquid cycle are not for use with patients.
† Sterilize time shown is the minimum validated sterilize time, and can only be adjusted to a longer time.
Figure 6-9. Liquid Cycle In-Cycle Screens
Cycle Selection ScreenCycles Screen
Condition Phase (Green) Sterilize Phase (Blue) Exhaust Phase (Violet)
Press for More Cycles
NOTE: Countdown timer on the display is estimated; timer self-corrects estimated time at the
beginning of each phase.