6-22
129390199 Operator Manual Sterilizer Operation
Table 6-10. Factory-Set Liquid Cycle and Cycle Values
The AMSCO Evolution Series Prevacuum Sterilizer is equipped with the following factory programmed liquid
sterilization cycle and cycle values.
Cycles Sterilize Temp.
Sterilize
Time
Dry
Time Recommended Load
Validation
Standard
LIQUID
*
250°F (121°C) 45 Min. N/A Three 1000 mL flasks.
R
EFER TO Table 6-5 FOR
GUIDELINES.
ST-8
* Liquids sterilized using the Liquid cycle are not for use with patients.
† Sterilize time shown is the minimum validated sterilize time, and can only be adjusted to a longer time.
Figure 6-9. Liquid Cycle In-Cycle Screens
Cycle Selection ScreenCycles Screen
Condition Phase (Green) Sterilize Phase (Blue) Exhaust Phase (Violet)
Press for More Cycles
NOTE: Countdown timer on the display is estimated; timer self-corrects estimated time at the
beginning of each phase.