6-15
Sterilizer Operation Operator Manual 129390199
6.9 Prevac Cycles Prevac cycles, 270°F (132°C), are used for sterilizing double-
wrapped instrument trays or fabric packs. Prevac cycles, 275°F
(135°C), are used for sterilizing double-wrapped instrument trays
only.
1. Before running this cycle refer to S
ECTION 6.1, BEFORE OPERATING
THE STERILIZER, on PAGE 6-1.
2. See instructions for using the loading car/transfer carriage or
rack and shelves, earlier in this section, for procedures to load
the sterilizer.
For 270°F (132°C) Cycle, press Prevac button to start the
prevacuum cycle. Refer to Table 6-1 and Table 6-6 for
descriptions of cycle use and loading recommendations.
For 275°F (135°C) Cycle press MORE CYLES touch screen
pad, then press PREVAC touch-screen pad to start the
275°F (135°C) Prevacuum cycle. Refer to Table 6-1 and
Table 6-6 for descriptions of cycle use and loading
recommendations.
NOTE: If the wrong cycle has been selected, see S
ECTION
6.15, ABORTING CYCLES, on PAGE 6-36 in this section.
3. Sterilizer automatically progresses through cycle, as follows:
NOTE: The control display indicates cycle progress in three
ways:
a. The large numerals in the screen’s center display estimated
remaining cycle time,
b. A status bar below the large numerals indicates approximate
cycle progress,
c. Screen background color changes as cycle progresses:
green indicating beginning (Condition), blue indicating middle
(Sterilize) and violet indicating end of cycle (Exhaust).
ACTIVATE SEAL Steam enters the door seal, pressing it
against inside surface of the door.
PURGE Chamber is purged with steam. Start of Condition is
printed.
VACUUM/PRESSURE PULSES
270°F (132°C) #1 - #7 Three vacuum pulses (drawing from
2 inHg to 24 inHg) are followed by three pressure pulses
(0.2 psig to 15 psig).
275°F (135°C) #1 – #3 – Vacuum point is printed and
pressure/vacuum pulse is repeated.
CHARGE Chamber is charged with steam. Start of steam
charge is printed.
STERILIZE Start of sterilize exposure is printed when the
chamber reaches sterilization temperature. Chamber
temperature is printed every minute.
FAST EXHAUST Start of exhaust is printed and chamber is
exhausted to 4 psig (0.28 bar).
WARNING – BURN HAZARD:
Sterilizer, rack/shelves and
loading equipment will be hot
after cycle is run. Always wear
protective gloves and apron
when removing a processed
load. Protective gloves and
apron must be worn when
reloading sterilizer following
the previous operation.
Steam may be released from
the chamber when door is
opened. Step back from the
sterilizer each time the door is
opened to minimize contact
with steam vapor.
WARNING SLIPPING
HAZARD: To reduce slippery
conditions, keep floors dry by
immediately wiping up any
spilled liquids or
condensation in sterilizer
loading or unloading area.
WARNING PERSONAL
INJURY and/or EQUIPMENT
DAMAGE HAZARD: When
closing the chamber door,
keep hands and arms out of
the door opening and make
sure opening is clear of
obstructions.
!
!
!