EasyManuals Logo

Amsco Evolution Series User Manual

Amsco Evolution Series
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #147 background imageLoading...
Page #147 background image
9-13
Troubleshooting Operator Manual 129390199
9.3.2 Too Long To
Open OE (or NOE)
Door
Occurs if door switch does
not open within the allotted
time.
Alarm Screen
Alarm Detail
Door switch did not open in
allotted time.
Causes and Correction:
1. Door switch malfunction
Check LS1/LS3 (LS2/LS4)
connections
Readjust LS1/LS3 (LS2/
LS4)
Repair LS1/LS3 (LS2/LS4)
2. Power door mechanism
failure
Repair mechanism
Readjust door cable
Replace motor
NOTE: Door switch may fail
closed to a closed state when
hot. Verify both 1 and 2
(above) then replace switch.
9.3.3 Pressure In
Chamber
Occurs if 2 psig (0.14 bar)
pressure is sensed in the
chamber.
Alarm Screen
Alarm Detail
2 psig (0.14 bar) pressure
sensed in chamber when not in
cycle.
Causes and Correction:
1. Valve Malfunction
•Repair V02
2. Control out of calibration
Recalibrate (contact
qualified Service person)
3. Transducer, PT1,
malfunction
Test transducer
Replace
Recalibrate
Table 9-2. Operator Troubleshooting Out-of-Cycle Alarms (Continued)
Condition Alarm Screen on Sterilizer Control Service Information

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Amsco Evolution Series and is the answer not in the manual?

Amsco Evolution Series Specifications

General IconGeneral
BrandAmsco
ModelEvolution Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals