6-7
Sterilizer Operation Operator Manual 129390199
6.4 Loading Cart and
Transfer Carriage
Instructions: Loading
1. Open sterilizer door.
2. Verify loading cart is properly aligned upon transfer carriage.
3. Align front end of transfer carriage with dock at base of sterilizer.
(See Figure 6-1 and Figure 6-2.)
4. Roll carriage forward until latches engage with dock.
5. Press forward on carriage handle, transfer carriage latches to
dock as cart begins pushing into chamber.
6. Carefully push cart into chamber.
7. Once cart is fully inserted, turn hand crank (See Figure 6-1 and
Figure 6-2) counterclockwise to lower carriage support arms
(disengaging the cart from carriage).
8. Pull carriage away from sterilizer chamber. Loading cart remains
in chamber.
9. Once transfer carriage is fully extracted from chamber, it
disengages from dock. Move carriage away from sterilizer.
10. Sterilizer is now ready to run a cycle. Proceed to cycle
description found in S
ECTION 6.9, PREVAC CYCLES, on PAGE 6-15.
Figure 6-1. Transfer Carriage and Loading Cart
Hand Crank
Transfer
Carriage
Loading Cart
Transfer
Carriage
Docking
Interface