EasyManua.ls Logo

Amsco Evolution Series - Control Measure for Verifying Sterilization Process; Biological Monitors; Testing for Prevacuum Efficiency; Dart (Bowie-Dick) Test

Amsco Evolution Series
176 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-2
129390199 Operator Manual Techniques of Sterilization
4.2 Control Measure
for Verifying
Sterilization Process
As part of the operator’s verification of the sterilization process,
biological indicators may be used to demonstrate that sterilization
conditions have been met.
4.2.1 Biological Monitors A live spore test utilizing G. stearothermophilus is the most reliable
form of biological monitoring. This type of product utilizes controlled
populations of a controlled resistance, so that survival time and kill
time can be demonstrated.
To verify the process, insert the biological indicator in a test pack
and place pack on the bottom shelf. Run test pack through a typical
cycle. On completion, forward test pack and monitor to appropriate
personnel for evaluation. Refer to AAMI guidelines to conduct routine
biological monitoring.
4.2.2 Testing for Prevacuum
Efficiency
Tests such as the Dart
1
(Bowie-Dick) are designed to document the
removal of residual air from a sample challenge load.
Run a Dart (Bowie-Dick) test cycle daily before processing any loads
in a sterilizer equipped with prevacuum cycles. The first prevacuum
cycle of each day should be used to test the adequacy of air removal
from the chamber and load, so that steam can penetrate the load. It
is not a test for adequate exposure to heat in terms of time-at-
temperature.
4.2.3 Dart (Bowie-Dick) Test
A steam penetration test should be carried out at the beginning of
each day the sterilizer is to be used. Refer to instructions for running
the Dart (Bowie-Dick) test cycle in S
ECTION 6.14.1, DART (BOWIE-
D
ICK) TEST ON PAGE 6-31. Dart (Bowie-Dick) test packs are designed
to expose the pattern and document the removal of residual air from
the sample load. Any test package must be constructed in
accordance with instructions given in the ANSI/AAMI ST-79
standard.
In the case of this test, following exposure in a prevacuum test cycle,
the pack is opened, the indicator examined and conclusions are
drawn as to the pattern of residual air, if any, that remained in the
pack during the sterilizing cycle. Any indication of a malfunction
must be reported to the supervisor. Sterilizer must not be used to run
prevacuum cycles until approved by supervisor. The cause of the
malfunction must be determined and resolved prior to operating the
sterilizer.
1. In this context and elsewhere, where implied (such as on screen buttons), DART is an acronym for Daily Air Removal Test. However, note
that the term is also registered as a testing apparatus, i.e., Dart
®
Testing Apparatus to Determine the Effectiveness of Removing Air from a
Chamber.
WARNING STERILITY
ASSURANCE HAZARD: Load
sterility may be compromised
if the biological indicator or
vacuum leak test indicates a
potential problem. If these
indicators show a potential
problem, refer the situation to
a qualified service technician
before using the sterilizer
further.
!
WARNING STERILITY
ASSURANCE HAZARD: Load
sterility may be compromised
if the biological indicator or
vacuum leak test indicates a
potential problem. If these
indicators show a potential
problem, refer the situation to
a qualified service technician
before using the sterilizer
further.
!

Table of Contents

Related product manuals