8-2
129390199 Operator Manual Routine Maintenance
8.2 Daily Maintenance Clean Chamber Drain Strainer
Important: The chamber drain strainer must be cleaned at least
once a day, preferably in the morning before running the first cycle.
1. Remove the drain strainer from the drain in the bottom of the
chamber.
2. Remove any obvious debris from the strainer. If necessary, clear
the screen in the strainer using a brush, wire or similar tool.
3. Once it has been cleared of obvious debris, reverse flush the
strainer under running water.
4. Replace the strainer in the chamber drain.
8.2.1 Replace Printer Paper
Roll
The printer paper roll should be checked every day. Replace the
paper roll whenever a colored stripe is visible on one or both edges
of the printout paper.
1. Open thermal printer front cover. Note cover is magnetically held
closed. (See Figure 8-2.)
Figure 8-2. Thermal Printer
WARNING – BURN HAZARD:
Allow sterilizer and
accessories to cool to room
temperature before
performing any cleaning or
maintenance procedures.
Figure 8-1. Remove and Clean Chamber Drain Strainer