7-15
Cycle and Control Value Programming Operator Manual 129390199
7. If an incorrect value is entered, press the backspace key to clear
the entry, then use the keypad to enter the correct value.
8. Once the correct year value has been entered, press OK to
accept.
NOTE: The day of the week is automatically understood and
registered by the control.
9. Press EXIT to return to Change Time and Date screen.
7.5 Change Machine
Setup
All changes are made to displayed settings using touch pads on the
control screen. No mechanical adjustments to the sterilizer are
necessary.
The Change Machine options are used to change the way the
sterilizer operates in a general way. The control has an User Login
security feature. All options are accessed from the Change Machine
screen.
A summary of the setup options that can be adjusted are listed in
Table 7-4. Each value is detailed in this section.
Table 7-4. Change Machine Setup
MACHINE SETUP
Utilities Control This setting permits the operator to program the sterilizer to automatically shut off its
steam and water at the end of the work day, to conserve utilities. Shut down and
power-up times can be programmed for any time of the day on Weekdays or the
Weekend. If the sterilizer is equipped with an integrated electric steam generator; an
automated generator flush can be programmed.
Language This option can be used to select one of three factory set languages as the default
for displays and printouts. Available languages are ENGLISH, SPANISH and
FRENCH.
Machine Number This is used to assign a two character, alphanumeric code to the sterilizer. This
code appears in the heading of all printouts.
Print Format This setting allows the sterilizer to print a condensed version of the cycle printout to
conserve on paper usage.
Audible Signals This option allows adjustment of ALARM, END OF CYCLE and TOUCH PAD
signals tones. Tones can be independently adjusted to one of three volume levels.
ALARM signal tone cannot be turned off.
Units This is used to select the temperature and pressure units displayed and printed out
by the sterilizer. The sterilizer is capable of displaying temperature as either
Fahrenheit or Celsius; pressure can be displayed as psig/inHg or bar.
Print Options Sterilizer can be set to automatically furnish a duplicate printout of each cycle at the
end of the cycle. First line will always read — DUPLICATE PRINT — and complete
printout of cycle data will be furnished.
Machine Access This option is used to set the Supervisor Password. Setting the Supervisor
Password prevents access to the Set Time and Date, Machine Setup and Change
Cycle Values functions on the Options screen.