9-6
129390199 Operator Manual Troubleshooting
9.2.4 Too Long In
Air Break
Occurs if chamber does not
air break the vacuum to 2
inHg (0.1 vbar) within the
allotted time.
Alarm Screen
Alarm Detail
Chamber did not air break
vacuum to 2 inHg (0.1 vbar)
within allotted time
Causes and Correction:
1. Air inlet filter plugged
• Replace
2. Valve malfunction
•Repair V01
3. Control out of calibration
• Recalibrate (contact
qualified Service person)
9.2.5 Under Sterilize
Temperature
Occurs if chamber
temperature drops below
sterilize temperature.
Alarm Screen
Alarm Detail
Chamber temperature
dropped below sterilize
temperature.
Causes And Correction:
1. Steam pressure
< 50 psig (< 3.4 bar)
• Check steam supply
piping
2. Valve malfunction
•Repair V02
3. Control out of calibration
• Recalibrate (contact
qualified Service person)
NOTE: Under sterilize
temperature value is set to
- 0°, and is not adjustable
Table 9-1. Operator Troubleshooting – In-Cycle Alarms (Continued)
Condition Alarm Screen on Sterilizer Control
Service Information