9-6
129390199 Operator Manual Troubleshooting
9.2.4 Too Long In
Air Break
Occurs if chamber does not
air break the vacuum to 2
inHg (0.1 vbar) within the
allotted time.
Alarm Screen
Alarm Detail
Chamber did not air break
vacuum to 2 inHg (0.1 vbar)
within allotted time
Causes and Correction:
1. Air inlet filter plugged
Replace
2. Valve malfunction
•Repair V01
3. Control out of calibration
Recalibrate (contact
qualified Service person)
9.2.5 Under Sterilize
Temperature
Occurs if chamber
temperature drops below
sterilize temperature.
Alarm Screen
Alarm Detail
Chamber temperature
dropped below sterilize
temperature.
Causes And Correction:
1. Steam pressure
< 50 psig (< 3.4 bar)
Check steam supply
piping
2. Valve malfunction
•Repair V02
3. Control out of calibration
Recalibrate (contact
qualified Service person)
NOTE: Under sterilize
temperature value is set to
- 0°, and is not adjustable
Table 9-1. Operator Troubleshooting In-Cycle Alarms (Continued)