Installation Verification Operator Manual 129390199
3.2.3 Utility Requirements •Electric:
208/240 V ac, 3-phase, 50/60 Hz, 12 A / phase,
480 V ac, 3-phase, 50/60 Hz, 6 A / phase
See Equipment Drawings for electrical requirements for sterilizers
with integral electric steam generators.
Steam:
Pressure: 50 to 80 psig (3.5 to 5.5 bar)
Consumption, Peak: Approximately 455 lbs/hr (206 kg/hr)
Water:
Pressure: 20 to 50 psig (1.4 to 3.5 bar)
Temperature: 40 - 70°F (4 - 21°C)
Consumption: 16 gpm (61 lpm), peak
•Air:
Pressure: 80 to 120 psig (5.5 to 8.3 bar)
Consumption: 3 cfm (5.1 cmh) max
3.2.4 Environmental
Conditions
Temperature: 50° to 90°F (10° to 32°C)
Humidity: 10 to 90% noncondensing
Pollution Degree: 2
Installation Category (Overvoltage Category): II
A-Weighted Sound Power Level: = 85 dBA (maximum)