Sterilizer Operation Operator Manual 129390199
6.10 Gravity Cycles The Gravity cycle is used for sterilizing double-wrapped instrument
trays and fabric packs.
1. Refer to S
ECTION 6.1, BEFORE OPERATING THE STERILIZER, on PAGE
6-1 before running this cycle.
2. See instructions for using the loading car/transfer carriage or
rack and shelves, earlier in this section, for procedures to load
the sterilizer.
3. To start the Gravity cycle, press the appropriate GRAVITY
touch-screen pad. Refer to Table 6-9 for descriptions of cycle
use.
4. Sterilizer automatically progresses through cycle, as follows:
NOTE: The control display indicates cycle progress in three
ways:
a. The large numerals in the screen’s center display estimated
remaining cycle time,
b. A status bar below the large numerals indicates approximate
cycle progress,
c. Screen background color changes as cycle progresses:
green indicating beginning (Condition), blue indicating middle
(Sterilize) and violet indicating end of cycle (Exhaust).
NOTE: If the wrong cycle has been selected, see S
ECTION
6.15, ABORTING CYCLES, on PAGE 6-36 in this section.
ACTIVATE SEAL – Steam enters the door seal, pressing it
against inside surface of the door.
PURGE – Chamber is purged with steam. Start of condition is
printed.
CHARGE – Chamber is charged with steam. Start of steam
charge is printed.
STERILIZE – Start of sterilize exposure is printed when the
chamber reaches sterilization temperature. Chamber
temperature is printed every minute.
FAST EXHAUST – Start of exhaust is printed and chamber is
exhausted to 4 psig.
DRY – Start of dry is printed and display counts down dry time
remaining.
AIR BREAK – Chamber is returned to atmospheric pressure.
RETRACT SEAL – A vacuum is drawn on the seal, retracting it
from inner surface of door.
COMPLETE – Complete tone sounds. Cycle summary and end of
cycle messages are printed.
5. Unload sterilizer (see instructions for using the loading
car/transfer carriage or rack and shelves, earlier in this section
for procedures to remove load)
WARNING BURN HAZARD:
Sterilizer, rack/shelves, and
loading equipment will be hot
after cycle is run. Always wear
protective gloves and apron
when removing a processed
load. Protective gloves and
apron must be worn when
reloading sterilizer following
the previous operation.
Steam may be released from
the chamber when door is
opened. Step back from the
sterilizer each time the door is
opened to minimize contact
with steam vapor.
WARNING SLIPPING
HAZARD: To reduce slippery
conditions, keep floors dry by
immediately wiping up any
spilled liquids or
condensation in sterilizer
loading or unloading area.
WARNING PERSONAL
INJURY HAZARD: When
closing the chamber door,
keep hands and arms out of
the door opening and make
sure opening is clear of
obstructions.
!
!
!