EasyManua.ls Logo

Amsco Evolution Series - Waste Temperature Probe Failure; RTC Failure

Amsco Evolution Series
176 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9-14
129390199 Operator Manual Troubleshooting
9.3.4 Waste
Temperature Probe
Failure
Occurs if waste line
temperature reading is
outside the normal range.
Alarm Screen
Alarm Detail
RTD probe, RTD4, output is
outside normal range.
Causes and Correction:
1. Loose connections in probe
wiring
•Repair
2. Probe failed
Replace
Recalibrate
3. Control out of calibration
Recalibrate (contact
qualified Service person)
9.3.5 RTC Failure
Occurs if real time clock
timer does not agree with
program clock.
Alarm Screen
Alarm Detail
The real time clock and the
application program clock are
not within 1% of each other.
Causes and Corrections:
1. Controller is defective
Replace controller.
Table 9-2. Operator Troubleshooting Out-of-Cycle Alarms (Continued)
Condition Alarm Screen on Sterilizer Control Service Information

Table of Contents

Related product manuals