Table 6-9. Factory-Set Gravity Cycles and Cycle Values
The AMSCO Evolution Series Prevacuum Sterilizer is equipped with the following factory programmed gravity
sterilization cycles and cycle values.
Cycles Sterilize Temp.
Sterilize
Time
Dry
Time Recommended Load
Validation
Standard
GRAVITY 250°F (121°C) 30 Min. 15 Min.
Fabric packs.
REFER TO Table 6-4 FOR
RECOMMENDED LOADS BY
STERILIZER CHAMBER SIZE.
ST-8
GRAVITY 270°F (132°C) 15 Min.
30 Min.
*
Double-wrapped instrument trays,
max. weight of 25 lbs (11.3 kg)
each.
REFER TO Table 6-4 FOR
RECOMMENDED LOADS BY
STERILIZER CHAMBER SIZE.
ST-8
GRAVITY 250°F (121°C) 30 Min. 30 Min.*
Double-wrapped instrument trays,
max. weight of 25 lbs (11.3 kg)
each.
REFER TO Table 6-4 FOR
RECOMMENDED LOADS BY
STERILIZER CHAMBER SIZE.
ST-8
GRAVITY 270°F (132°C) 25 Min. 15 Min.
Fabric packs.
R
EFER TO Table 6-4 FOR
RECOMMENDED LOADS BY
S
TERILIZER CHAMBER SIZE.
ST-8
* Dry time for Floorloading Sterilizers is 45 minutes.
† Sterilize times shown are the minimum validated sterilize times, and can only be adjusted to longer times..
Figure 6-7. Gravity Cycle In-Cycle Screens
Cycle Selection Screen
Condition Phase (Green) Sterilize Phase (Blue) Exhaust Phase (Violet)
Press for More Cycles
NOTE: Countdown timer on the display is estimated; timer self-corrects estimated time at the
beginning of each phase.