9-21
Troubleshooting Operator Manual 129390199
9.4.10 OE/NOE
Door Switches
Malfunction
Occurs if control senses
OE or NOE door is in the
open and closed position
simultaneously.
(Applicable to sliding-
door sterilizers only.)
Alarm Screen
Alarm Detail
Door switches do not agree.
Causes and Correction:
1. Door switch malfunction
Check LS1(LS3) connections
for OE door or LS2 (LS4) For
NOE Door
Readjust door switches
Determine which switch is in
the incorrect state, and
replace that switch
9.4.11 Exhaust
Rate Too Slow
Occurs if liquid cycle slow
exhaust rate is too slow.
Chamber exhausted slower than
the expected rate.
Causes and Correction:
1. Solenoid valve malfunction
Repair V40
2. Solenoid valve malfunction
Repair V03
3. Control out of calibration
Recalibrate (contact qualified
service person)
Table 9-3. Operator Troubleshooting Sensor Alarms (Continued)