EasyManua.ls Logo

Amsco Evolution Series - Test Cycles; Table 6-12. Test Cycles

Amsco Evolution Series
176 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6-30
129390199 Operator Manual Sterilizer Operation
6.14 Test Cycles Test cycles are factory programmed on prevacuum sterilizers. These
cycles are used to verify the sterilizer is functioning at optimum
capability.
Table 6-12. Test Cycles
Test Cycles for
All Units
Sterilize
Temp. Sterilize Time
Dry
Time Recommended Load
Validation
Standard
Dart Warm-up
*
270°F (132°C) 4 Min. 1 Min. N/A N/A
Dart (Bowie-Dick)
Test
*
270°F (132°C) 3 Min., 30 Sec 1 Min. Dart or Bowie-Dick Test Pack ST-8
Leak Test
*
270°F (132°C) N/A N/A N/A ST-8
* Not adjustable.
Figure 6-13. Test Cycles Start - Cycle Screens
Cycles Screen Test Cycles Screen
Press the right arrow button at the lower right hand
corner of the screen to advance to Test Cycles screen.
Press the desired test cycle button (Warm-Up Cycle,
Dart Test, or Leak Test) to run the required cycle.
ACCESSING TEST CYCLES

Table of Contents

Related product manuals