3-2
129390199 Operator Manual Installation Verification
Air Supply:
Supply line shutoff with provisions for lockout and tagout located
near the equipment.
Air pressure 80 to 120 psi (5.5 to 8.3 bar), dynamic.
3.1.3 Electrical Service The protective earth ground must be connected to terminal block
TB-1 in the sterilizer power box.
Three-phase power for the sterilizer must meet specifications on
the Equipment Drawing.
Verify proper rotation of the vacuum pump by observing pump
rotor shaft.
Important: Three-phase service requires a clearly marked
disconnect with lockout/tagout capability located near the
sterilizer with easy access for maintenance personnel.
3.1.4 Sterilizer Final Check Chamber leveled properly.
Door opens and closes smoothly.
Door locked switches adjusted correctly.
Chamber strainer in place.
Rack and shelves and/or loading cart operate correctly.
Paper loaded in printer.
Warranty labels properly applied.
Verify programming and set up support proper configuration of
unit.
3.1.5 Cycle Operation Unit powers up correctly.
Run Warm Up Cycle
Run Leak Test cycle leak rate is to be less than 1.0 mm Hg/minute
(1.3 mbar/min).
Verify operation of a typical cycle (270°F [132°C] prevacuum).