6-26
129390199 Operator Manual Sterilizer Operation
Table 6-11. Factory-Set Cycles and Cycle Values SFPP Sterilizer
Cycles Sterilize Temp.
Sterilize
Time
†
Dry
Time Recommended Load
Validation
Standard
SFPP 270°F (132°C) 4 Min.
30 Min.
*
Double-wrapped instrument
trays, max. weight of 25 lbs
(11.3 kg) each. Fabric
packs.
R
EFER TO Table 6-4 FOR
R
ECOMMENDED LOADS BY
S
TERILIZER CHAMBER SIZE.
ST-8
SFPP 275°F (135°C) 3 Min. 30 Min.* Double-wrapped instrument
trays, max. weight of 25 lbs
(11.3 kg) each.
R
EFER TO Table 6-4 FOR
R
ECOMMENDED LOADS BY
S
TERILIZER CHAMBER SIZE.
ST-8
SFPP 270°F (132°C) 4 Min. 5 Min. Single fabric pack. ST-8
* Dry time for Floorloading Sterilizers is 45 minutes.
† Sterilize times shown are the minimum validated sterilize times, and can only be adjusted to longer times.
Figure 6-11. SFPP Cycle In-Cycle Screens
Cycle Selection ScreenCycles Screen
Condition Phase (Green) Sterilize Phase (Blue) Exhaust Phase (Violet)
Press for More Cycles
NOTE: Countdown timer on the display is estimated; timer self-corrects estimated time at the
beginning of each phase.