9-19
Troubleshooting Operator Manual 129390199
9.4.6 Chamber
Temperature Probe
Failure
Occurs if chamber
temperature reading
outside the normal range.
Alarm Screen
Alarm Detail
RTD probe, RTD1, output is
outside normal range.
Causes and Correction:
1. Loose connection in probe
wiring
Repair
2. Probe failed
Replace
Recalibrate
3. Control out of calibration
Recalibrate (contact qualified
Service person)
9.4.7 Jacket
Temperature Probe
Failure
Occurs if jacket
temperature reading is
outside the normal range.
Alarm Screen
Alarm Detail
RTD probe, RTD3, output is
outside normal range.
Causes and Correction:
1. Loose connection in probe
wiring
Repair
2. Probe failed
Replace
Recalibrate
3. Control out of calibration
Recalibrate (contact qualified
Service person)
Table 9-3. Operator Troubleshooting Sensor Alarms (Continued)