EasyManuals Logo

Amsco Evolution Series User Manual

Amsco Evolution Series
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #169 background imageLoading...
Page #169 background image
10-5
Service Procedures Operator Manual 129390199
Figure 10-2. Door Seal Replacement, 26 x 37.5” (660 x 953 mm) Sterilizers
Use flat tool with
rounded edges
(such as a non-
serrated table
knife) to pry and
twist one section of
the seal partially
from the groove.
Hinged Door Model Shown
Seal
Door
Chamber
Opening
The following lot data is molded into the back side of
the seal:
Location
of Seal
Reference
Indicators
Back Side
of Door Seal
Reference Indicator,
Enlarged
P129373-376 ABCDEFHGIJKLMNOP 1234 07 P
Part Number Lot Index Cure Year Manufacturer
Quarter Code
The lot index and cure quarter are struck off during manufacturing.
The first characters in sequence
not struck
are those used for
identification. For example:
ABCDEFHGIJKLMNOP 1234 07 P
The cure date is: 3rd Quarter, 2007. The lot is: 02C.