iii
Introduction Operator Manual 129390199
1.
Table 2. Factory-Set Cycles and Cycle Values for Evolution SFPP Sterilizer
Cycles Sterilize Temp.
Sterilize
Time
Dry
Time Recommended Load
Validation
Standard
1. SFPP 270°F (132°C) 4 Min. 30 Min.* Double-wrapped instrument trays,
max. weight of 25 lbs (11.3 kg) each.
Fabric packs.
R
EFER TO Table 4 FOR
R
ECOMMENDED LOADS BY STERILIZER
C
HAMBER SIZE.
ST-8
2. SFPP 275°F (135°C) 3 Min.
30 Min.
*
Double-wrapped instrument trays,
max. weight of 25 lbs (11.3 kg) each.
R
EFER TO Table 4 FOR
R
ECOMMENDED LOADS BY STERILIZER
C
HAMBER SIZE.
ST-8
3. PREVAC 270°F (132°C) 4 Min. 30 Min. Double-wrapped instrument trays,
max. weight of 25 lbs (11.3 kg) each.
Fabric packs.
R
EFER TO Table 4 FOR
R
ECOMMENDED LOADS BY STERILIZER
C
HAMBER SIZE.
ST-8
4. GRAVITY 250°F (121°C) 30 Min. 15 Min. Fabric packs.
R
EFER TO Table 4 FOR
R
ECOMMENDED LOADS BY STERILIZER
C
HAMBER SIZE.
ST-8
5. SFPP 270°F (132°C) 4 Min. 5 Min. Single fabric pack. ST-8
6. PREVAC 275°F (135°C) 3 Min. 30 Min. Double-wrapped instrument trays,
max. weight of 25 lbs (11.3 kg) each.
R
EFER TO Table 4 FOR
R
ECOMMENDED LOADS BY STERILIZER
C
HAMBER SIZE.
ST-8
7. GRAVITY 270°F (132°C) 15 Min. 30 Min.* Double-wrapped instrument trays,
max. weight of 25 lbs (11.3 kg) each.
R
EFER TO Table 4 FOR
R
ECOMMENDED LOADS BY STERILIZER
C
HAMBER SIZE.
ST-8
8. LIQUID
**
250°F (121°C) 45 Min. N/A Three 1000 mL flasks.
R
EFER TO Table 5 FOR GUIDELINES.
ST-8
* Dry time for Floorloading Sterilizers is 45 minutes.
** Liquids sterilized using the Liquid cycle are not for use with patients.