129390199 Operator Manual Sterilizer Operation
printed every minute. Chamber is controlled at set point plus
overdrive.
FAST EXHAUST Start of exhaust is printed and chamber is
exhausted to 4 psig (0.28 bar).
DRY Start of dry is printed and display counts down dry time
remaining.
AIR BREAK Chamber is returned to atmospheric pressure.
RETRACT SEAL A vacuum is drawn on the seal, retracting it
from inner surface of door.
COMPLETE Complete tone sounds. Cycle summary and end of
cycle messages are printed.
6. Once the cycle is complete:
a. Open the chamber door.
b. Unload the Bowie-Dick test pack.
c. Forward the exposed Bowie-Dick indicator to the
appropriate personnel for examination.
Important: Any indication of a malfunction must be reported to the
supervisor. Sterilizer must not be used again until approved by
supervisor. The cause of the malfunction must be determined and
resolved prior to operating the sterilizer.
WARNING – BURN HAZARD:
Sterilizer, rack/shelves and
loading equipment will be hot
after cycle is run. Always wear
protective gloves and apron
when removing a processed
load. Protective gloves and
apron must be worn when
reloading sterilizer following
the previous operation.
Steam may be released from
the chamber when door is
opened. Step back from the
sterilizer each time the door is
opened to minimize contact
with steam vapor.
!