Section 4 Operation
4-128
1
Main Func
On Off
1
RX Parameter
MT8801 99-12-31 12:00:00
<< Setup Analog RX Measure Parameter (CDMA) >>
->
Back
Screen
User Cal Factor
AF Input
Impedance
: [100k‰]
: [ 30V]
: [ 0.00dB]
Range
Fig. 4-24 Setup Analog RX Measure Parameter Screen
• Set the following items:
Item Range Resolution Initial value
User Cal Factor –55.00 dB to 55.00 dB 0.01 dB 0.00 dB
Note 1
AF Input
Range 30 V, 4 V, 400 mV, 40 mV 30 V
Impedance 100 kΩ, 600 Ω 100 kΩ
Note 1:
<User Cal Factor>
When a loss is between the RF output of this machine and RF input of the
mobile station, set a User Cal Factor; the RF level of the signal generator can
be converted to the input level of the mobile station.
For example, when a loss (5 dB) is between the RF output of this machine and
RF input of the mobile station, set 5 dB to the User Cal Factor.
• Main function key: None
• Function keys:
[Back Screen] F12: Displays the previous screen