Section 4 Operation
4-132
RF Level function keys:
1st page
[Incremental Step Value] F7 Sets the step value for up/down the RF level by the [Step] keys.
Range: 0.1 dB Set value 80.0 dB, 0.1 dB step
Initial value: 1.0 dB
[Relative On Off] F8 Displays the relative value with the reference value that is the set value when this key
is pressed.
Initial value: Off
When the level is set with numeric keys in Relative On mode, it becomes the actual
output level (not relative value).
Relative displayed value = Set value by numeric keys - value when this key is pressed
[Unit EMF TERM] F10 Selects the RF level unit of open voltage (EMF, dBµ) or termination voltage (TERM,
dBµ).
Selectable only when in dBµ display mode. 30 dBµ EMF = 24 dBµ TERM
Initial value: EMF
[Back Screen] F12 Returns to the Setup Common Parameter screen.
2nd page
[Offset Value] F7: Sets an offset value in output level offset display mode.
Range: 55.0 to 55.0 dB
Resolution: 0.1 dB
Initial value: 0 dB
[Offset On Off] F8: Increases or decreases the real output level to display.
Displayed value = Real output level value + Offset value
Range: On, Off
Initial value: Off
Note:
If a level is set using a numeric key when this item is On, the value becomes the
displayed value.
Real output level value = Displayed value - Offset value
[Back Screen] F12: Displays the previous screen.