4-131
4.3 CDMA Transmitter and Receiver Test --- TX and RX Tester Mode
2nd page
[Range Up] F7 Up the measurement range of the AF level meter.
[Range Down] F8 Down the measurement range of the AF level meter.
[Storage Mode] F9 Displays the Storage Mode menu for all the measured results on the screen.
[Normal] F7 Sets normal mode. (Initial value)
[Average] F8 Sets average mode.
[Average Count] F9 Sets number of Averaging processings.
2 ≤ Set value ≤ 9999
Initial value: 10
(In the average mode, the measurement is of single mode, which displays the
averaged results in each measurement, and stops measurement when the Aver-
age Count reached.)
Note that the Power Meter has not the average mode.
[return] F12 Returns to the AF Level menu.
[AF Level Unit] F10 Selects the unit of the AF Level measurement value of dBm (valid for 600 Ω of input
impedance) or V.
Initial value: V
When the 100 kΩ of Impedance of AF Level Input is set on the Setup Common Pa-
rameter (Analog) screen, this menu is not displayed.
[Distortion Unit] F11 Selects the unit of the distortion measurement value of dB or %.
Initial value: %
[Back Screen] F12 Returns to the Setup Common Parameter screen.
• RF Frequency function keys:
[Incremental Step Value] F7 Sets the step value for up/down the RF frequency by the [Step] keys.
1 Hz ≤ Set value≤ 3 GHz, 1 Hz step
Initial value: 1 MHz
[Relative On Off] F8 Displays the relative value with the reference value that is the set value when this key
is pressed.
Initial value: Off
When the frequency is set with numeric keys in Relative On mode, it becomes the
actual output frequency (not relative value).
Relative displayed value = Set value by numeric keys - value when this key is pressed
[Channel] F9 Changes the channel number. (See para. 4.4 for the changing method.)
[Back Screen] F12 Returns to the Setup Common Parameter screen.