Section 7 Calibration
7-2
7.1 Calibration Requirements
Calibration is done to help maintain the MT8801C’s performance.
Calibration should be performed periodically even if the MT8801C is operating nor-
mally.
We recommend that the MT8801C be calibrated once or twice a year.
Contact the Service Department of Anritsu Corporation if the MT8801C fails to meet
the specifications during calibration.
7.2 Equipment Required for Calibration
The table below shows the equipment required to calibrate each item.
Table 7-1 Equipment Required for Calibration
Recommended equipment Required performance† Calibration item
Frequency counter (MF1603A)
Frequency standard
• 100 KHz to 3 GHz
• Resolution: 1 Hz
• External reference input: 10 MHz
Standard radio-wave receiver or
equipment having equivalent function
(accuracy better than 1 × 10
–9
)
Frequency accuracy of reference
Frequency accuracy of reference
crystal oscillator
Note:
Extracts part of performance which can cover the measurement range of the
test item.