1.
In the navigation pane of the Data Collection Software, select GA
Instruments4 ga37304 Plate Manager.
2.
At the boom of the screen, click New.
3.
Enter plate record information, then select Spectral Calibration in the
Application dropdown list.
4.
In row A01, enter a sample name, click the next cell, then enter a comment if
needed.
5.
In column 1 of row A01, select the instrument protocol y
ou created for the
spectral calibration.
Create a spectral
calibration plate
Chapter 8 Calibrate the instrument
Perform spectral calibration for sequencing and fragment analysis
8
100
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide