Monthly maintenance
Task
Frequency
Run the Water Wash wizard (see “Typical conditions for using maintenance
wizards“ on page 117).
Flush the array port during the Water Wash wizard, whether or not bubbles are
present in the array port.
Monthly or as needed
Check disk space (see “Check available disk space on E: drive“ on page 131 ). Monthly
As-Needed maintenance
Task
Frequency
Clean the drip tray. As needed
Change the capillary array using the Install Array wizard (see “Typical conditions
for using maintenance wizards“ on page 117).
As needed
Remove any dried polymer from the capillary tips. Use a lint-free wipe moistened
with deionized water.
As needed
Chapter 9 Maintenance
Maintenance schedules
9
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide
115