6.
If the array knob and ferrule are on the array tip, remove them and rinse them
with deionized water. Dry the parts with a lab wipe.
7.
Clean the array tip carefully with a lab wipe moistened with deionized water.
8.
Aach the array tip shipping vial lled with 1X run buer to the array tip.
Loosen the vial cap slightly, insert the tip and then tighten the cap.
9.
Clip the vial with the array tip onto the array frame.
10.
Store the capillary array upright in a safe area.
IMPORTANT! Check the 1X run buer lev
els in the shipping cover and vial at
least once a week; replenish the buer as necessary to keep both ends of the
capillaries immersed in buer.
Clean the detection cell
1.
Open the instrument door and the ov
en door.
2.
Open the detection cell door by turning the door knob to the left, then lowering
the door.
2 3
1
1
Detection cell door knob
2
Capillary array knob
3
Capillary array
3.
Turn the capillary array knob 1/4-turn clockwise. The pointer on the knob is
pointed to the left.
4.
Pull the pump and lower polymer blocks forward until the detection cell is out of
the detection block.
Chapter 9 Maint
enance
Clean the detection cell
9
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide
125

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 3730xl and is the answer not in the manual?

Applied Biosystems 3730xl Specifications

General IconGeneral
BrandApplied Biosystems
Model3730xl
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals