5.
Hold an empty tube or beaker under the exit ing
to receive approximately 5 mL of waste. Flush the
trap by pushing steadily on the syringe plunger.
IMPORTANT! DO NOT USE EXCESSIVE FORCE.
Take approximately 30 seconds to ush 5 mL of
either distilled or deionized water through the trap.
6.
Close the ings by turning each clockwise until the
ings seal against the block. Close the Luer ing
rst, then close the Exit ing.
IMPORTANT! Do not ov
er-tighten the ings. The ings require only enough
tightening to prevent water leaks. Excessive tightening can damage the ings.
7.
Remov
e the syringe from the Luer ing. Hold the ing with one hand while
turning the syringe counterclockwise with the other hand.
Install the capillary array
• Capillary array, 96- or 48-capillary
• Lab wipes, lint-free
• Gloves
IMPORTANT! W
ear gloves when you handle the capillary array.
Use the Install Array wizard when changing capillary arrays to properly ll the
capillaries with polymer.
1.
In the navigation pane, select GA Instruments4 ga37304 Instrument
name.
2.
At the top of the screen, select W
izards4Install Array wizard.
3.
Follow the instructions in the wizard to install the capillary array.
IMPORTANT! When this wizard runs, it introduces polymer into the anode
buer jar
. Replace the anode jar buer after you run this wizard. See “Fill the
anode buer jar“ on page 33.
4.
P
erform a spatial calibration (see “Perform spatial calibration“ on page 91).
Required
mat
erials
Install a new or
used capillary
array
Chapter 9 Maint
enance
Install the capillary array
9
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide